Metrohm 840 PC Control 6.0 / Touch Control User Manual

Page 13

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

1 General

■■■■■■■■

7

Range

1 ... 14

Default value

1

Name

The designation of the favorite is used for unambiguous identification.

Input

24 characters maximum

Type

Definition, whether the favorite is representing a single method or a com-
plete sample data silo.

Selection

Method | Sample data silo

Default value

Method

Memory

Memory location the method or the sample data silo is loaded from. Only
the currently accessible memory locations are selectable.

Selection

Internal memory | Card 1 | Card 2 | Shared
memory

Default value

Internal memory

Shared memory
Memory released on an external system. For this an 847 USB Lab Link
is necessary.

Method / Silo

Method or sample data silo of the favorite.

Selection

Selection of stored methods or sample data
silos

[More options]

Parameterize the autostart function, see following chapter.

More options
With [More options] the autostart function is parameterized.

Autostart

on | off (Default value: off)

If this parameter is activated, a new determination is started automatically
at the end of a determination. This continues until the quantity specified
has been reached (see Number of autostarts).

Number of autostarts

This parameter can only be edited when Autostart = on.

Number of automatic starts.