5 cell, 6 control device, Cell – Metrohm 840 PC Control 6.0 / Touch Control User Manual

Page 21: Control device

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

2 Parameters

■■■■■■■■

15

2.1.5

Cell

Under [Cell], the parameters for the titration cell are defined.

Generator electrode

Type of generator electrode.

Selection

without diaphragm | with diaphragm

Default value

without diaphragm

without diaphragm
Generator electrode without diaphragm.
with diaphragm
Generator electrode with diaphragm.

Generator current

Polarization current at the generator electrode.

Selection

100 mA | 200 mA | 400 mA | auto

400 mA
Default value, if Generator electrode = without diaphragm.
auto
The current is adapted to the conductivity of the reagent and automat-
ically reduced near the endpoint. Default value, if Generator elec-
trode
= with diaphragm.

Reagent monitoring

Selection of the reagent from the list of reagents. The selection depends
on the titration mode. Reagents are defined under System

Reagents.

Selection

Selection of configured reagents | off

Default value

off

off
The reagent monitoring is not active.

2.1.6

Control device

Under [Control device], the control device is selected the determination
is carried out with. Control devices are defined under System

Device

manager.

Note

This button is only displayed when several control devices have been
configured.

Advertising