3 stop conditions, 4 conditioning, Stop conditions – Metrohm 840 PC Control 6.0 / Touch Control User Manual

Page 18: Conditioning

Advertising
background image

2.1 Coulometric Karl Fischer titrations (KFC)

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

(see Chapter 2.1.3, page 12). Entering an extraction time may e.g. be
advisable if a Karl Fischer oven is used.

Range

0 ... 999999 s

Default value

0 s

Temperature

Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement
= automatic or continuous is defined (see sensor dialog
of the command).

Range

–20.0 ... 150.0 °C

Default value

25.0 °C

Time interval MP

Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.

Range

0.1 ... 999999.0 s

Default value

2.0 s

2.1.3

Stop conditions

Under [Stop conditions], the conditions for stopping a titration are
defined, if this does not occur automatically. This could be the case when
the endpoint set is not reached or if the stop criterion (see "Stop crite-
rion", page 10)
is not fulfilled.

Stop time

The titration is stopped when the specified time has elapsed since the start
of the titration.

Range

1 ... 999999 s

Selection

off

Default value

off

2.1.4

Conditioning

Under [Conditioning], the conditions required for conditioning are
defined.

Conditioning

on | off (Default value: on)

If this parameter is activated, then the first time the titration is started the
working medium will be titrated to the endpoint with the specified control
parameters. The status is kept stable. The actual method run does not
begin until [START] has been pressed once more.