7 sensor, Sensor – Metrohm 840 PC Control 6.0 / Touch Control User Manual

Page 22

Advertising
background image

2.1 Coulometric Karl Fischer titrations (KFC)

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

Control device

Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command. The control devi-
ces do not have to be connected. This allows to transfer methods more
easily from one system to another.

Selection

Selection of configured control devices

2.1.7

Sensor

Under [Sensor], the parameters for the sensor are edited.

Sensor

Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System

Sensors. You can

also enter a sensor name which is not contained in the sensor list. When a
determination is started there is a check whether the sensor is contained
in the sensor list.

Selection

Selection of configured sensors

I(pol)

The polarization current is the current that is applied to a polarizable elec-
trode during the voltametric measurement.

Selection

5 µA | 10 µA | 20 µA | 30 µA

Default value

10 µA

Electrode test

on | off (Default value: off)

For polarizable electrodes and the generator electrode, an electrode test
can be carried out. The following will be tested:

Polarizable electrodes

– Is the electrode connected?
– Is there a short circuit?

The electrode test is carried out when this command is started.

Generator electrode:

– Is the electrode connected?
– Is the generator system defective?
– Is the resistance of the sample solution very high?

The electrode test is carried out continuously during the execution of
this command.

Temp. measurement

Type of temperature measurement.