Metrohm 840 PC Control 6.0 / Touch Control User Manual
Page 9
■■■■■■■■■■■■■■■■■■■■■■
1 General
■■■■■■■■
3
■
at the start of the determination.
■
at the end of the determination.
Number of determ.
The number of determinations to be carried out with a certain amount of
reagent depends on the type of sample and its amount.
Range
1 ... 999
Selection
off
Default value
off
Working life
Working life of the reagent.
Range
1 ... 999 days
Selection
off
Default value
off
Volume
This parameter is only visible with volumetric reagents.
Volume of the titrant dosed.
Range
1.0 ... 999.9 mL
Selection
off
Default value
off
Reagent capacity
This parameter is only visible with coulometric reagents.
Water capacity of the reagent.
Range
1 ... 9999 mg
Selection
off
Default value
off
Drift
This parameter is only visible with coulometric reagents.
During conditioning the measured drift has to be in the following range
for 2 min: 'specified drift value + 50 µg/min'.
Range
1 ... 999 µg/min
Selection
off
Default value
off
[Status]
Display the status overview of the current values of the reagent monitor-
ing.