3 stop conditions, 4 conditioning, Stop conditions – Metrohm 840 PC Control 6.0 / Touch Control User Manual

Page 27: Conditioning

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

2 Parameters

■■■■■■■■

21

Extraction time

Minimum duration of the titration. The titration will not be stopped during
the extraction time, even if the endpoint has already been reached. The
titration is however stopped if a stop condition is fulfilled during this time
(see Chapter 2.2.3, page 21).

Range

0 ... 999999 s

Default value

0 s

Temperature

Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement
= automatic or continuous is defined (see sensor dialog
of the command).

Range

–20.0 ... 150.0 °C

Default value

25.0 °C

Time interval MP

Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.

Range

0.1 ... 999999.0 s

Default value

2.0 s

2.2.3

Stop conditions

Under [Stop conditions], the conditions for stopping a titration are
defined, if this does not occur automatically. This could be the case when
the endpoint set is not reached or if the stop criterion (see "Stop crite-
rion", page 19)
is not fulfilled.

Stop time

The titration is stopped when the specified time has elapsed since the start
of the titration.

Range

1 ... 999999 s

Selection

off

Default value

off

2.2.4

Conditioning

Under [Conditioning], the conditions required for conditioning are
defined.