2 operation via «ic net – Metrohm 762 IC Interface User Manual
Page 26
3 Operation
762 IC Interface
22
3.2
Operation via «IC Net»
This section describes only the most important points concerning the
operation of the 762 IC Interface. For further details please refer to the
«IC Net» Instructions for Use and to the on-line help in the PC pro-
gram.
3.2.1
Settings in the "762 IC Interface" window
By clicking the 762 icon in the toolbar with the left mouse button or by
clicking this icon with the right mouse button and selecting the
Open
menu item the
762 IC Interface
window for parameter settings is opened.
It consists of three tabs
Init
,
RS settings
, and
Links
.
Init
The
Init
tab of the
762 IC Interface
window contains data acquisition pa-
rameters for the 762 IC Interface.
Label
Optional label to
name the interface with
maximum 8 characters.
ROM version
Instrument program version number.
Measuring per second
Number of data points measured per s.
Entry range:
10, 20, 30, 50, 60 points/s
Range
Range for AD converter .
Example: converts
±
2500 mV to 2
24
bits
Entry range:
2500, 1250, 625, 312.5, 156.25, 78.125,
39.062 mV