Metrohm 762 IC Interface User Manual
Page 34
4 Appendix
762 IC Interface
30
FIFO
0
…
16 Bytes (for Dev. 8 not available)
Pin assignment
Pin 2:
RxD (Received Data) for Dev. 1
…
4
Pin 3:
TxD (Transmitted Data) for Dev. 1
…
4
Pin 5:
GND (Signal Ground) for Dev. 1
…
4
Pin 7:
TxD (Transmitted Data) for Dev. 5
…
8
Pin 8:
RxD (Received Data) for Dev. 5
…
8
Pin 9:
GND (Signal Ground) for Dev. 5
…
8
Remote interface (Events)
Output lines
7 potential-free relay contacts per system for
controlling external instruments.
RUN
COM
Maximum contact load:
30 V / 1 A DC
Input lines
1 digital input per system for external method
start.
+5V
470k
47k
Ext. Start
RUN
COM
Maximum load for external contact:
approx. 0.1 mA at max. cut-off voltage of approx.
5 V
Analog signal interface
Function
2 potential-free analog signal inputs per system.
Connector
Dsub connector 9 pin (female)
1
5
6
9
Input voltage range
±
2.5 V
Input amplification
1, 2, 4, 8, 16, 32, 64 (selectable)
Sampling rate
10, 20, 30, 50, 60 measuring points/s
Resolution
24 bit (1 LSP = 0.298
µ
V at amplification 1)
Noise
< 20
µ
Vpp
< 3
µ
Vrms (at 0 V input voltage, amplification 1
and sampling rate 10 Hz)