Metrohm 762 IC Interface User Manual

Page 34

Advertising
background image

4 Appendix

762 IC Interface

30

FIFO

0

16 Bytes (for Dev. 8 not available)

Pin assignment

Pin 2:

RxD (Received Data) for Dev. 1

4

Pin 3:

TxD (Transmitted Data) for Dev. 1

4

Pin 5:

GND (Signal Ground) for Dev. 1

4

Pin 7:

TxD (Transmitted Data) for Dev. 5

8

Pin 8:

RxD (Received Data) for Dev. 5

8

Pin 9:

GND (Signal Ground) for Dev. 5

8

Remote interface (Events)

Output lines

7 potential-free relay contacts per system for
controlling external instruments.

RUN

COM

Maximum contact load:
30 V / 1 A DC

Input lines

1 digital input per system for external method
start.

+5V

470k

47k

Ext. Start

RUN

COM

Maximum load for external contact:
approx. 0.1 mA at max. cut-off voltage of approx.
5 V

Analog signal interface

Function

2 potential-free analog signal inputs per system.

Connector

Dsub connector 9 pin (female)

1

5

6

9

Input voltage range

±

2.5 V

Input amplification

1, 2, 4, 8, 16, 32, 64 (selectable)

Sampling rate

10, 20, 30, 50, 60 measuring points/s

Resolution

24 bit (1 LSP = 0.298

µ

V at amplification 1)

Noise

< 20

µ

Vpp

< 3

µ

Vrms (at 0 V input voltage, amplification 1

and sampling rate 10 Hz)

Advertising