3 Operation

762 IC Interface

26

Program

On the

Program

subpage, program steps including time, program in-

struction and parameter can be entered.

First column

Time at which program instruction is applied.
Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruction is
applied together with the last instruction with
time entry.

Second column

Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be en-
tered if activated on the

Events configuration

tab.

Third column

Parameter for program instruction (see below).

ENABLED

Enable program start (a disabled program is
not started).

<

Add

>

Add new program instruction.

<

Delete

>

Delete selected program instruction.

<

Verify

>

Test the time program (error messages are dis-
played if program is wrong).

List of program instructions

The following program instructions can be added to the time program

on the

Program

subpage:

Instruction

Entry

Meaning

Events

0, 1, p, *

Set

event output lines 1...7 to the desired

values. For entry of the first value, enter

1

,

0

,

p

or

*

. For entry of the other values, move

the cursor in front of the value to be
changed and enter

1

,

0

,

p

or

*

.

Program

END, RESET

The

END

flag can be used to end a

program, especially if the program time
should be longer than the chromatogram
duration. Additional steps after this flag are
not allowed. The

RESET

flag is used to

reset the parameters to the system startup
values.

Advertising