Metrohm 762 IC Interface User Manual

Page 35

Advertising
background image

4.1 Technical data

762 IC Interface

31

Zero error

±

2.5 mV

Pin assignment

Pin 1,4,5,8,9: GND (ground, connected to

housing)

Pin 2:

Analog 1+

Pin 3:

Analog 1–

Pin 6:

Analog 2+

Pin 7:

Analog 2–

Safety specifications

Construction/testing

According to IEC 1010 / EN 61010 / UL 3101-1,

protection class 1, degree of protection IP40

Safety directions

The Instructions for Use include information and

warnings which must be heeded by the user to

assure safe operation of the instrument.

Electromagnetic compatibility (EMC)

Emitted interference

Standards met:

EN55011 (level B), EN55022 (level B),

EN 50081-1/2, EN61000-3-2

Immunity to interference

Standards met:

IEC61000-4-2/EN61000-4-2 (level 3),

IEC61000-4-3/EN61000-4-3 (level 3),

IEC61000-4-4/EN61000-4-4 (level 4),

IEC61000-4-5/EN61000-4-5 (level 2/3),

IEC61000-4-6/EN61000-4-6 (level 3),

IEC61000-4-11/EN61000-4-11,

IEC61000-4-14/EN61000-4-14 (level 3),

EN50082-2, NAMUR

Ambient temperature

Nominal operating range

+5

+45°C

(at 20

80 % atmospheric humidity)

Storage, transport

–40

+70°C

Housing

Material of cover

Polyurethane rigid foam (PUR) with fire protection

for fire class UL94VO, CFC-free

Material of base

Steel, enameled

Dimensions

Width

255 mm

Height

128 mm

Depth

340 mm

Weight (with accessories)

2.762.0010: 4.3 kg

2.762.0020: 4.9 kg