9troubleshooting – Trebor Q-Series Di Water Heater User Manual

Page 30

Advertising
background image

PAGE 30

Q-SERIES DI HEATER OPERATION / MAINTENANCE MANUAL

9

TROUBLESHOOTING

Display is Not Illuminated

Cause:

Solution:

No power at main circuit breaker

Review wiring procedure (see Section 3, Installation).

Control breaker is blown

Check breaker.

Main breaker is OFF

Turn Main breaker to ON.

Alarm Sounds

Cause:

Solution:

DI supply pressure low

Check water supply pressure at source.

Liquid level low

Check water supply at source.

Check overpressure relief diaphragm for rupture.

Check fluid lines in and out of manifold for flow.

Liquid leak

Check manifold and heater modules for leak. Visually
inspect leak tray.

Thermocouple damaged

Check thermocouple leads at temperature limit controller
for continuity.

Over-temperature alarm

Check if heater element has failed and replace module
(see section 8.1).

Check thermocouple leads at temperature limit controller
for continuity.

Verify fan operation.

Unable to maintain set point

Cause:

Solution:

Failed heater element

Circuit breaker tripped

Check if heater element has failed and replace module
(see Section 8.1).

Check and reset the main, auxiliary and element isolation
circuit breakers.

Consult Trebor if circuit breaker repeatedly trips.

Solid State Relay (SSR) may have
“frozen” open

An SSR has failed in the open mode: Shut system off
immediately and replace the failed SSR. SSR’s should be
replaced in pairs.