ITC ProTime Microcoagulation System User Manual

Page 12

Advertising
background image

12

Touch the large drop of blood to the collection cup. Keep adding blood until the blood level

fills the cup above the line.

For Tenderlett Plus LV, ensure the cup is filled completely. Ensure the blood extends all
the way to the bottom of the cup. Add another drop if you are not sure you have

enough.

5. Snap Tenderlett Plus to ProTime

Hold the device at an angle and place the front end of the device into the slot in the
instrument.

Press down to click the Tenderlett Plus in place. You should hear a soft click.

Note: Proper engagement of the
Tenderlett Plus to the cuvette is
critical to prevent a sample error.

6. Start the Test

Press the button to start the test. This signals ProTime to draw the sample into the
cuvette.