Troubleshooting Guidelines

Become skilled at obtaining blood with the fingerstick technique. This will help avoid sample errors.

Follow all of the directions on the screen for optimal operation.

Do not try to hasten the test process by performing the fingerstick while the WARMING screen is
displayed, as blood may clot before it is drawn into the cuvette and an error may occur. FOLLOW THE

INSTRUCTIONS ON THE SCREEN.

Troubleshooting Guide

Screen Display

Source of Error

Method of Control

INR HIGH
REPEAT TEST
IF SAME, CALL MD

Patient’s PT result is too
high (INR >12.0)

Repeat the test. If it occurs again,
patient should be referred to their
physician IMMEDIATELY and re-tested
at the lab.

INR LOW
REPEAT TEST
IF SAME, CALL MD

Patient’s PT result is too low
(INR < 0.6)

Repeat the test. If it occurs again,
patient should be referred to their
physician IMMEDIATELY and re-tested
at the lab.

ON-BOARD QC
OUT OF RANGE
TRY AGAIN

Level I or II control is too
high or too low

Repeat the test. Check for adequate
storage of cuvettes or sampling
technique. Confirm results with the
lab.

NO CLOT DETECTED
REPEAT TEST
IF SAME CALL MD

At least one channel did not
clot

Repeat the test. If it occurs again,
patient should be referred to their
physician IMMEDIATELY and re-tested
at the lab.

TIME OUT
TRY AGAIN

Time ran out before starting
test

Test again with new Tenderlett Plus
and a fresh fingerstick. You may reuse
the cuvette if no blood was in contact
with the cuvette.

TEST INTERRUPTED
TRY AGAIN

Cuvette removed or operator
interrupted test

Review correct procedure. Repeat test.

SAMPLE NOT SEEN
TURN OFF
TRY AGAIN

Can be caused by small
samples, clots, air bubbles,
or a seal problem in one or
more channels

Check for proper collection technique.
Verify cup is completely filled. Use
another cuvette.

SAMPLE TOO LARGE
TURN OFF
TRY AGAIN

Sample oversize

Review directions and repeat the test.

SAMPLE TOO SMALL
TURN OFF
TRY AGAIN

Sample undersize

Repeat test. Exceed the fill line on the
cup to ensure sample size.

SAMPLE ERROR
TURN OFF
TRY AGAIN

Incorrect fluid movement
Air bubble detected in one
or more of the channels

Turn off and on again and repeat test
with new cuvette and fresh fingerstick.