Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

The ProTime® Microcoagulation System is intended for use in the electromagnetic environment specified
below. The customer or the user of the ProTime® Microcoagulation System should assure that it is used in
such an environment.

Immunity Test

IEC 60601 Test
Level

Compliance
Level

Electromagnetic Environment -
Guidance

Electrostatic
Discharge (ESD)

IEC 61000-4-2

+ 6 kV contact
+ 8 kV air

+ 6 kV contact
+ 8 kV air

Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.

Electrical fast
transient / Burst

IEC 61000-4-4

+ 2 kV for
power supply
lines
+1 kV for
input/output
lines

+ 2 kV for
power supply
lines
+1 kV for
input/output
lines

Mains power quality should be that of a
typical domestic, commercial or hospital
environment.

Surge
IEC 6100-4-5

+1 kV
differential mode
+2 kV common
mode

+1 kV
differential mode
+2 kV common
mode

Mains power quality should be that of a
typical domestic, commercial or hospital
environment.

Voltage dips,
short
interruptions,
and voltage
variations on
power supply
input lines

IEC 61000-4-11

<5 % U

T

(>95 % dip in

U

T

for 0.5 cycle)

40 % U

T

(60 % dip in U

T

)

for 5 cycles
70 % U

T

(30 % dip in U

T

)

for 25 cycles
<5 % U

T

(>95 % dip in

U

T

) for 5 sec

<5 % U

T

(>95 % dip in

U

T

for 0.5 cycle)

40 % U

T

(60 % dip in U

T

)

for 5 cycles
70 % U

T

(30 % dip in U

T

)

for 25 cycles
<5 % U

T

(>95 % dip in

U

T

) for 5 sec

Mains power quality should be that of a
typical domestic, commercial or hospital
environment.

If the user of the ProTime

®

Microcoagulation System requires
continued operations during power mains
interruptions, it is recommended that the
ProTime

®

Microcoagulation System be

powered from an uninterruptible power
supply or battery.

Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical domestic, commercial
or hospital environment.

NOTE: U

T

is the AC mains voltage prior to application of the test level.