ITC ProTime Microcoagulation System User Manual

Page 24

Advertising
background image

24

SET PID/OID
The PID/OID selection is indicated only by the location of the highlight (text on the screen is unchanged).

Use the button to select PID/OID options:

Selecting PID/OID ON enables both a patient ID and operator ID to be entered.

Selecting PID ON enables only a patient ID to be entered.

Selecting OID ON enables only an operator ID to be entered.

Selecting OFF disables both a patient ID and operator ID.
Note: The selection is highlighted and remains the default until changed again by the user.

Entering and Changing Numerical Information (PID/OID, PASSWORD)
By using a combination of the button and button and the triangular cursor on the display screen,
numeric information can be entered into the instrument for operator identification number (OID), patient
identification number (PID) and PASSWORD fields. When entering numeric information into a field,
pressing a button has the following actions:

CHANGE - Pressing the button once will increment the digit in the current cursor position by

one, starting at 0 and going through 9 and back to 0 again.

OK - Pressing the button briefly enters the digit displayed in the current cursor position and then

moves the cursor right, to the next field position. Pressing the button for 2 seconds or longer saves
the complete numeric field.

EXAMPLE: Entering a PID number:
With the PID feature enabled, the following screen appears after a cuvette is inserted into the instrument: