Main menu options – ITC ProTime Microcoagulation System User Manual

Page 16

Advertising
background image

16

Screen Display

Source of Error

Method of Control

CUVETTE EXPIRED

Expired cuvette

Check to make sure cuvettes have not
expired. The expiration date is located
on the packaging and alongside the
barcode on the cuvette.

BARCODE ERROR
REMOVE CUVETTE
TRY AGAIN

Instrument cannot read
barcode accurately

Visually inspect barcode. If scratched,
discard cuvette. If dirty, wipe clean. If
barcode is good, review correct
procedure and repeat test.

BATTERY ERROR
TURN OFF
TRY AGAIN

Power supply battery error

Repeat the test. If it occurs again,
replace the battery. If problem
persists, call ITC Technical Support.

INSTRUMENT ERROR
TURN OFF
TRY AGAIN

Instrument set up, data log
or communication error

Repeat the test. If problem persists,
call ITC Technical Support.

CHARGE ERROR
TURN OFF
TRY AGAIN

Power supply error

Check the AC/DC power module,
repeat the test. If problem persists,
call ITC Technical Support.

TEMP ERROR
TURN OFF
TRY AGAIN

Temperature not in range

Check for proper operating
temperature. Repeat the test. If
problem persists, call ITC Technical
Support.

PHOTO ERROR
TURN OFF
TRY AGAIN

LED blocked or other photo
system error

Repeat the test. If problem persists,
call ITC Technical Support.

MAIN MENU OPTIONS