ITC ProTime Microcoagulation System User Manual

Page 13

Advertising
background image

13

It takes only a few seconds for ProTime to draw the blood into the cuvette. Watch the
screen for the next prompt.

7. Remove Tenderlett Plus

Remove Tenderlett Plus immediately when prompted to do so.
CAUTION: Failure to do so will result in an error message. ProTime allows you six
seconds.

Do not press the button after the Tenderlett Plus is removed from the ProTime
instrument while you are testing your INR. This will interrupt the test procedure, and you

will have to start over with a new blood sample.

The instrument then progresses to the test and displays the TESTING screen.