2.

3

Error messages

E1

When the

scale

is switched on, the weight on the

platform is above or greater than the automatic zero
range (greater than 10% of the capacity of the scale).
Make sure that there is no weight on the platform. If the
error persists, re-calibrate the machine.

E2

When the

scale

is switched on, the weight on the

platform is below or lower than the automatic zero
range (greater than 10% of the capacity of the scale).
Correct service set up procedure has not been
followed. Ensure calibration weight is applied and
removed exactly in accordance with instructions.

E3

Not used.

E4

The weight on the platform is unsteady.

E5

There is a problem with the load cell. Check the wiring
(see page

60

). If the error persists, replace the load cell

(see page

33

).

E6

The internal zero value is greater than 360,000. The
error may occur in calibration mode.

E7

The internal zero value is less than 240,000. The error
may occur in calibration mode.

E8

This message will only appear for

legal for trade

scales

.

If you are configuring the capacity of the

scale

, the

number of divisions cannot exceed 10,000.

The weight is greater than the capacity of the

scale

.

Temporary error. Remove any weight from the platform,
switch off and then on again. If this error persists,
contact your authori

z

ed service agent.