5 Configuration Mode

5.6 F3 - zero display setting

5.6

F3 - zero display setting

MUST BE SET TO 0 AT

LEGAL FOR TRADE SCALES.

If the load cell gives unstable readings set to 5 which is
the maximum.

If the load cell gives stable readings set to 0 which is the
minimum.

Ensure the hold function (F8) is set to hold 0.

Zero display setting
0 =

default

zero display reading within

1 division
2 = zero display reading within
2 divisions

3 = zero display reading within
3 divisions
4 = zero display reading within
4 divisions
5 = zero display reading within
5 divisions

To change the value:

3 times

F3

ENTER

2Ero0

2Ero1

ENTER

F3