4.3 Setup Menu Parameters

This section provides more detailed descriptions of the selections found in the Setup
Menu.

The menu table shows the options and default parameter in LCD display format to
coinside with the actual display.

Table 4.2 Parameter Choices and Explanations

SubMenu1

Option

Default

Parameter Description

Comment

P1

00-15

05

Auto off time

00 = no auto off
01-15 = auto off time in minutes. The scale will
automatically turn off after specified time when
there has been no flucuation of weight.

p2

0-2

2

Hold key function

0 = only HOLD function
1 = only PRINT function
2 = HOLD and PRINT function. Short press for
Print function and long press for Hold function.

p3

0-2

2

Hold mode function

0 = no hold function
1 = hold larger weight reading
2 = auto release hold function when weigh is
below 10d and auto hold new stable weigh
(more than 10d)

3-50

3-50 = unchangeable reading when the variety
is within ±3-50d

p4

0-7

2

Output of print data

0 = no RS232 function. Will not transmit or
receive data.
1 = output display data when PRINT is pressed
and weight reading is stable.
<LF><weight reading, minus sign, decimal
point, unit><CR><EXT>
2 = output gross, tare and net weight and weigh
unit when PRINT pressed.
<LF><Gross:weight reading, minus sign,
decimal point, unit><CR><EXT>
<Tare:weight reading, decimal point,
unit><CR><EXT>
<Net:weight reading, minus sign, decimal point,
unit><CR><EXT>
3 = continuously output displayed weight
reading and unit. No data will be received.
Output format same as choice 1.
4 = continuously output gross, tare, net weight
and weigh unit. No data will be received. Output
formats same as choice 2.
5 = output display data and weigh unit one time
when scale is stable. Output format same as
choice 1.
6 = output gross, tare, net weight and weigh
unit one time when scale become stable.
Output formats same as choice 2.
7 = Bio-RS232 data output, compatible to NCI-
SP1