2 display symbols, 3 error messages and troubleshooting, Display symbols error messages and troubleshooting – Salter Brecknell PS2000 User Manual

Page 36

Advertising
background image

34

PS1000/2000 User Instructions

7.2 Display Symbols

7.3 Error Messages and Troubleshooting

Symbol

Description

0------

Zero is over the setting range

0_______

Zero point is below the setting range

Ad------

Signal to ADC is over maximum range

Ad______

Signal to ADC is below minimum range

--------

Weight is over upper limitation or display data is over limitation

________

Weight is below lower limitation

eep.e0

EEPROM can’t be accessed.

EEP.E1

The parameters are not the same as the backup data

EEP.E2

The setting parameter(s) is not in normal range

Lo.bAt

Battery voltage is below 4.7V

CAP.- -

The setting full capacity will be displayed

CAL.Px

Calibration on point (x)

CAL.Er

Calibration error, maybe input data or loaded weight is incorrect, unstable, non-linear

d.--

The division will be displayed

px.y

The x = parameter, y = parameter choice

Symptom

Probable Cause

Remedy

Ad------

Loadcell wires to indicator are incorrectly
connected, shorted, opened, ADC or loadcell(s)
are damaged.

Make sure wires are ok and correctly connected.
Replace loadcell or ADC chip, Service required.

Ad______

0------

Weight reading exceeds Power On Zero limit.

Make sure scale platform is empty. Perform zero
calibration.

0______

Weight reading below Power On Zero limit.

Install platform on scale. Perform zero calibration.

--------

Weight reading exceeds Overload limit or the
weight value cannot be displayed in the current
unit of measure because it exceeds 6 digits.

Reduce load on scale until weight value can be
displayed. Use a more appropriate unit of measure.

________

Weight reading below under load limit.

Install platform on scale. Perform zero calibration.

EEP.E2

Parameter is not correctly set.

Re-set parameter

CAL.Er

Calibration error. Input data or loaded weight is
too small, too big, unstable, non-linear.

Input correct data, load correct weight onto
platform, Service required.

Will not turn on.

Power cord not plugged in or properly
connected. Power outlet not supplying
electricity. Battery discharged. Other failure.

Check power cord connections. Make sure power
cord is plugged into the power outlet. Check power
source. Replace batteries. Service required.

Cannot zero the display
or will not zero when
turned on.

Load on scale exceeds allowable limits. Load
on scale is not stable. Load cell damage.

Remove load on scale. Wait for load to become
stable. Service required.