PS1000/2000 User Instructions

17

P5

0-4

3

Baud rate for RS-232

0 = 1200 bps
1 = 2400 bps
2 = 4800 bps
3 = 9600 bps
4 = 19200 bps

p6

0-2

0

RS-232 protocol

0 = 8NO
1 = 7O1
2 = 7E1

p7

00-31

9

Scale resolution select

Refer to

Table 4.3

for resolution choices and

values.

P8

0-2

1

Calibration division size

Refer to

Table 4.4

and

Table 4.5

0 = 1
1 = 2
2 = 5

p9

0-5

1

Decimal point in calibration Refer to

Table 4.4

and

Table 4.5

0 = 1
1 = 0.1
2 = 0.01
3 = 0.001
4 = 0.0001
5 = 10

p10

0,1

1

Calibration unit

Refer to

Table 4.4

and

Table 4.5

0 = kg
1 = lb

p11

0-6

6

Weigh units enable

0 = only kg
1 = only lb
2 = only lb:oz
3 = kg or lb
4 = kg or lb:oz
5 = lb or lb:oz
6 = kg, lb, or lb:oz

p12

0-7

3

Power on zero range

0 = calibration zero -point +1%FS
1 = calibration zero -point +2%FS
2 = calibration zero-point +5%FS
3 = calibration zero-point +10%FS
4 = calibration zero-point +20%FS
5 = calibration zero-point +50%FS
6 = calibration zero-point +100%FS
7 = no limitation

p13

0-7

2

Zero range for [ZERO] key 0 = power-on zero-point +1%FS

1 = power-on zero-point +2FS
2 = power-on zero-point +5%FS
3 = power-on zero-point +10%FS
4 = power-on zero-point +20%FS
5 = power-on zero-point +50%FS
6 = power-on zero-point +100%FS
7 = no limitation

p14

0-2

2

Weight signal within power
on zero point range

0 = current weight
1 = calibration zero-point
2 = switch off zero point and power off tare
weight as current tare weight

SubMenu1

Option

Default

Parameter Description

Comment