HI-4453/FP5083/FP4083/FP2083

© ETS-LINDGREN L.P. – June 2005

Rev. F, P# H-600064

32

When it completes the command, the probe responds with a
string consisting of

1) A start character (":")
2) The command letter, followed by
3) Data, if required, and terminated with
4) A carriage return.

If the command does not require the probe to return any data,
the probe simply responds with the command letter and a
carriage return. If an error occurs, the probe responds with an
error code, as detailed in Appendix A.

COMMANDS

Command

Description

Axxx

Axis enable/disable. x = "E" means enable, x =
"D" means disable. xxx order is X axis, Y axis, Z
axis

B

Read battery voltage.

Cx

Set baud rate.

x = 1 sets rate to 2400 baud

x = 2 sets rate to 9600 baud

New baud rate does not take effect until the next
power-up.

Dx

Read probe data.

x = 1 enables short form output

x = 2 enables long form output

Lx

Load table data.

x = ASCII hex data

Rx Set

range.

x = 1, 2, 3, 4 or N (next range)

Sx Sleep

timer.

x = number of seconds to wait for a command
before putting the probe into the sleep mode.

Tx Read

Temperature.

x = C or F

Ux

Set unit type.

x = 1, 2, 3, or N (next unit)

1 = V/m

2 = mW/cm²

3 = [V/m]²