Isotropic Electric Field Probes
HI-4453/FP5083/FP4083/FP2083

© ETS-LINDGREN L.P. – June 2005
Rev. F, P# H-600064

33

Command

Description

Vtx

Verify table data.

t = table (same table as for the Lx command,
above). x = number of bytes to send in one
string.

Z Zero.
Null

Send the ASCII null character. This is a special
command that can be used as the initial
command to the probe after it is powered up.
The probe responds with “N”.

Probe Output

Command

HI-4453/FP5083/FP4083/FP2083 Response

B

Bxx.xx, where xx.xx is the battery voltage.

D1

Dxx.xxuuu, the short form output.

xx.xx is the reading. The position of the decimal
point depends upon the range setting of the
probe.

uuu = units

_V_ = V/m, mW2 = mW/cm², _V2 = [V/m]²
(underscore indicates a space character)

D2

Dxx.xxuuurrrobaaat, the long form output.

xx.xx = the reading, as described for D1.

uuu = units, as describe for D1.

rrr = recorder out value (A 3-digit ASCII number
from 0 to 255).

o = over range indicator ("N" = ok, "O" = over
range).

b = battery status ("N" = safe operating level, "W"
= warning level, "F" = fail level).

aaa = axis enable ("E" = enabled).

t = terminating carriage return.

Rx

Rx, where x is the range.

x = "" returns the range currently in use

x = 1, 2, 3, 4 enables the selected range

x = N sets the probe to the next (higher) range.

TF

Txxx, where xxx is temperature in ° Fahrenheit.

TC

Txxx, where xxx is temperature in ° Centigrade.

Vtx

Vxx[xx...], where xx is table data in ASCII hex.

Archived 4/2/10