6 rs-232 installation instructions – Rice Lake Counting Scales - IQ6500 Installation and Operation Manual User Manual

Page 31

Advertising
background image

28

9.2.6

RS-232 Installation Instructions

Use the following procedure to install an RS-232 connection on your IQ6500.

1. Remove platter and four screws at center of platter support. Remove platter support.

2. Remove case cover by removing six screws on top and two screws under front edge of scale.

3. Unplug connectors (A and B).

4. Remove two screws (1) from each A/D board. Lift each board out and gently place them on top of

load cell.

5. Noting placement of ground wires, remove three screws (2) from A/D mounting bracket. Remove

mounting bracket.

6. Noting placement of ground wire, remove three screws (3)

from back panel.

7. Carefully push out RS-232 port covers from rear

panel.

8. The middle and right side brackets (4)

have metal spacers under them.
Remove these spacers and
mount RS-232 board (5 )
as shown under the
brackets.

9. Route RS-232 cable as

shown to main board CPU
in connector CN2. Use
wire ties to secure cables
safely away from case
edges and load cell
compartment. This
connector is keyed and
will only fit one way.

10. Reassemble scale being

careful not to pinch any
wires between case cover
and base.

11. Set specification codes to

activate appropriate
output ports.

To A

To B

A

B

Ground To

Load Cell

Mount

C

To Load

Cell

To C or

Second Load

Cell

RS-232

Cable to CN2

Connector

on CPU

1

1

A/D Boar

d

A/D Boar

d

A/D Mounting

Bracket

2

2

3

4

5