IQ6200 Battery Pack Installation

1. Gently lift and remove platform.

2. Remove four screws at center of platform support. Remove platform support.

3. Remove six screws from top covers (two in keypad section and four in main).

4. Remove two screws from front underside of scale (one in each corner) and remove main cover.

5. Remove two screws at top of each battery support (1). Note placement of ground wires (2) for rein-

stallation.

6. Place battery pack in compartment, as shown, with wires at top of battery towards rear of scale.

7. Replace supports (1) and ground wires (2). Do not allow wires to be pinched under supports.

8. Route wires as shown and plug connector (3) into board TPB-1677.

!

Caution

Plugging battery in with connector reversed can cause permanent damage to battery pack and scale.

9. Verify that scale (AC adaptor) is unplugged. Turn scale on to check battery operation.

If scale operates properly, reassemble scale and use accordingly.

If scale does not operate, recheck wiring and try again. If scale still does not operate, plug in
scale. If scale works with the power cord plugged in, turn scale off, reassemble scale, and allow
battery to charge (8 to 12 hours is recommended).

If scale does not operate with power cord, disconnect cord and battery and call authorized factory
service representative.