normally approximately 20,000 counts. To compute the number of counts divide the set

capacity by the increment size.

F6.3 Enter Cal Mode

This is how you calibrate the scale. Enable this and continue on to F6.4. If you do not

enable it you will advance to CAL d (calibration done) and the calibration will be

unchanged and you will exit from set-up mode.

F6.4 Linearity

Allows you to correct for small linearity errors in the middle of the weight range. This is

not allowed in “Legal for Trade” applications. If not used, this should be set to “0” if set

to “-0”, linearity will not be affected but you will be required to place “full capacity”

weight for calibration.

Set-Up Procedure

Place the internal CAL switch (S1-1) to the ON position. The scale must be configured

before operation by programming the needed parameters. Use the following front panel

switches to view and/or change existing data values:

[ZERO] Used to scroll backward through the programming sequence steps.

[TARE] Used to scroll forward through the programming sequence steps when in

function select mode and to sequence through available choices of individual functions.

[PRINT] Used to accept displayed data then to advance one step in the programming

sequence.

[CLEAR] Used to return to the Function Select Enable portion of the SET-UP procedure.

[F1] Used to advance to the end of the SET-UP procedure to save or discard the changes

entered. This button is not labeled and is located below the [RECALL TARE] button.

FSE Function Select Enable Selects the starting function group to be programmed. Use

[ZERO] and [TARE] buttons to move back and forth through this menu. Press print to

enter that section at the first item.

Choices: F1, F2, F3, F4, F5, F6 (F7 on some custom systems)

FSE F1

F1.0 Raw Counts Display Enable

d - Show normal counts on display.

Page 26