F6.3 Enter CAL Mode

d - Continue to last step (CAL d)

E - Enter CAL mode.

F6.4 Linearity Factor

Corrects non-linearity in weigh system. The value required is determined

empirically. Amount of correction depends upon percent usage of the load cell as

well as the system capacity and resolution.

Choices: 7 to -7

E SCALE Empty Scale

Remove all weight from the scale and then press the [PRINT] button.

888888 Full Scale Capacity

888888 indicates the previously entered Full Scale Capacity in even units. Use the

numeric keyboard to enter the Full Scale capacity. Press the [ENTER] button to

advance to the next step.

15 SEC Zero Setting Fifteen Second Delay

Counts down from 15 to zero to capture the zero setting. If the [SELECT] button is

pressed during the time-out period, the meter will immediately terminate the delay

and ignore the motion criterion to capture the weight. Allow at least three (3)

seconds after weight has been applied to the scale before pressing [SELECT] or the

weight captured may be erroneous.

Add Ld Add Load or Add ALL

The weight applied must be at least 5% of the full scale capacity. The message Add

ALL will appear if Linearity Factor is not zero indicating that Full Scale Capacity

must be applied.