F6.3 Enter CAL Mode
d - Continue to last step (CAL d)
E - Enter CAL mode.
F6.4 Linearity Factor
Corrects non-linearity in weigh system. The value required is determined
empirically. Amount of correction depends upon percent usage of the load cell as
well as the system capacity and resolution.
Choices: 7 to -7
E SCALE Empty Scale
Remove all weight from the scale and then press the [PRINT] button.
888888 Full Scale Capacity
888888 indicates the previously entered Full Scale Capacity in even units. Use the
numeric keyboard to enter the Full Scale capacity. Press the [ENTER] button to
advance to the next step.
15 SEC Zero Setting Fifteen Second Delay
Counts down from 15 to zero to capture the zero setting. If the [SELECT] button is
pressed during the time-out period, the meter will immediately terminate the delay
and ignore the motion criterion to capture the weight. Allow at least three (3)
seconds after weight has been applied to the scale before pressing [SELECT] or the
weight captured may be erroneous.
Add Ld Add Load or Add ALL
The weight applied must be at least 5% of the full scale capacity. The message Add
ALL will appear if Linearity Factor is not zero indicating that Full Scale Capacity
must be applied.