Rice Lake DWM-IV Weighmeter Standard Operators Manual User Manual

Page 39

Advertising
background image

[PRINT] - Advances to next step if span must be set.

[TARE] - Establishes a new zero calibration setting without changing the span

calibration previously set. The display will show CAL d. This feature is disabled if

the Add ALL message is displayed.

888888 Test Weight

Apply known weight to scale. Fractional or decimal weights are not acceptable—

only whole numbers. Use the numeric keyboard to enter the applied weight. Press

the [ENTER] button to advance to the next step. If Add ALL was displayed in the

previous step, the system will use the Full Scale Capacity as the test weight value

and automatically proceed to the next step.

15 SEC Span Setting Fifteen Second Delay

Counts down from 15 to zero to capture the span setting. If the [SELECT] button is

pressed during the time-out period, the meter will immediately terminate the delay

and ignore the motion criterion to capture the weight. Allow at least three (3)

seconds after weight has been applied to the scale before pressing [SELECT] or the

weight captured may be erroneous.

CAL d Calibration Done

Indicates that the calibration procedure is complete.

[PRINT] - Advances to next step.

[7] - Decreases span setting.

[9] - Increases span setting.

FCd File Changed Data

Changes will be saved or discarded.

0 - Save the entered values in memory and use again on power up.

1 - Discard present values and restore previously entered values.

2 - Initialize the EEPROM (U14) to the factory settings.

[PRINT] - Advance to the next step.