Series – Rice Lake Ranger Count 3000 Series, Ohaus User Manual
Page 16
EN-14
Ranger™ Count 3000
Series
Zero2
[ZErO]
Initiates a 2
nd
scale zero calibration.
Span1
[SpaN]
Initiates a span calibration procedure (zero and span). A span calibration is important when
initially setting up the scale.
Span2
[SpaN2]
Initiates a 2
nd
scale span calibration procedure.
Lin1 [LIN]
Initiates a linearity calibration procedure (zero, mid-point and span).
Lin2 [LIN2]
Initiates a 2
nd
scale linearity calibration procedure.
GEO [GEO]
Geographical Adjustment Factor (GEO) is used to adjust the calibration based on the current
location. Settings from 0 to 31 are available with 12 being the default. Refer to table 4-2 to
determine the GEO factor that corresponds to your location.
End Cal [End]
Advance to the next menu or return to the top of the current menu.
Zero2:
Perform
Span1:
Perform
Span2:
Perform
Linearity1:
Perform
Linearity2:
Perform
Geographic
Adjustment:
Set 0…12…31*
End Calibration: Exit menu
*Bold always represents factory default value.