Series – Rice Lake Ranger Count 3000 Series, Ohaus User Manual

Page 16

Advertising
background image

EN-14

Ranger™ Count 3000

Series














Zero2

[ZErO]

Initiates a 2

nd

scale zero calibration.

Span1

[SpaN]

Initiates a span calibration procedure (zero and span). A span calibration is important when

initially setting up the scale.

Span2

[SpaN2]

Initiates a 2

nd

scale span calibration procedure.

Lin1 [LIN]

Initiates a linearity calibration procedure (zero, mid-point and span).

Lin2 [LIN2]

Initiates a 2

nd

scale linearity calibration procedure.

GEO [GEO]

Geographical Adjustment Factor (GEO) is used to adjust the calibration based on the current

location. Settings from 0 to 31 are available with 12 being the default. Refer to table 4-2 to

determine the GEO factor that corresponds to your location.

End Cal [End]

Advance to the next menu or return to the top of the current menu.















Zero2:

Perform

Span1:

Perform

Span2:

Perform

Linearity1:

Perform

Linearity2:

Perform

Geographic

Adjustment:

Set 0…12…31*

End Calibration: Exit menu

*Bold always represents factory default value.