Rice Lake Ranger Count 3000 Series, Ohaus User Manual

Page 9

Advertising
background image

Ranger™ Count 3000 Series EN-7


3.2 Turning Scale On/Off

To turn the scale on, press and hold the On/Zero Off button for 1 second. The scale performs a

display test, momentarily displays the software version, and then enters the active weighing

mode.

To turn the scale off, press and hold the On/Zero Off button until OFF is displayed.


3.3 Counting

Ranger Count 3000 has three display areas. Weight, Piece Weight and Count information are

displayed in these areas respectively.

3.3.1 Preset tare

Enter a value with the numeric keypad. The value will be displayed in the Count window.

Pressing Tare key, the value in current unit will then be saved as the preset tare.

To clear the tare value, clear the pan and press Tare key.

3.3.2 APW Establishment

If there is no APW established, both the piece weight display and the count display will show 0.


Positive Sampling:

1. Place the sample on the pan.
2. Input the sample size with the numeric keypad. The value is displayed in the count

window.
Alternatively the Sample key can be used instead. Pressing the Sample key and

[------] is displayed in the count window. Then input the sample size value (only

integer).

3. Press the Sample key, [------] will be displayed in both the piece window and the count

window.

4. The established APW and count will be displayed.

Negative sampling:

1. Place container with the samples on the pan.
2. Tare the scale, a NET 0 will be displayed.
3. Remove the samples from the container; weight window will now display a negative net

reading.

4. Input the sample size with the numeric keypad. The value will be displayed in the count

window.
Alternatively the Sample key can be used instead. Pressing the Sample key and

[------] is displayed in the count window. Then input the sample size value (only

integer).

5. Press the Sample key, [------] will be displayed in both the piece window and the count

window.

6. The established APW and count will be displayed.

Advertising