6. MAINTENANCE

6.1 Cleaning

The housing may be cleaned with a cloth dampened with a mild detergent if necessary. Do not

use solvents, chemicals, alcohol, ammonia or abrasives to clean the housing or control panel.

6.2 Troubleshooting

Table 6-1 lists common problems and possible causes and remedies.

If the problem persists, contact OHAUS or your authorized dealer.

TABLE 6-1

Symptom

Possible Cause

Remedy

Cannot turn on

No power to scale

Verify connections and voltage

Poor accuracy

Improper calibration

Unstable environment

Perform calibration

Move scale to suitable location

Cannot calibrate

Unstable environment

Incorrect calibration weight

Move the scale to suitable location

Use correct calibration weight

Cannot access mode

Mode not enabled

Enter menu and enable mode

Cannot access unit

Unit not enabled

Enter menu and enable unit

Battery icon flashing

Battery is empty

Connect the scale to AC power

and charge the battery

Err 8.1

Power On Error

Weight reading exceeds Power On

Zero limit

Err 8.2

Power On Error

Weight reading below Power On

Zero limit

Err 8.3

Over Range Error

Weight reading exceeds Overload

limit

Err 8.4

Under Range Error

Weight reading below Underload

limit

Err 8.5

Tare out of range

Adjust tare value to be within

range

Err 8.6

Display overflow

Weight exceeds 6 digits

Err 9.5

Calibration data error

Calibration data not present

-------

Busy

Displayed during tare setting, zero

setting, printing

--NO--

Action not allowed

Function not executed

CAL E

Calibration error

Unstable environment

Incorrect calibration weight

Calibration value outside allowed

limits

Move the scale to suitable location

Use correct calibration weight

Lo.REF

Low reference weight warning

Increase reference weight.

rEF.Err

Unacceptable reference weight

Reference weight too small.

Weight on the pan is too small to

define a valid reference weight.

Increase reference weight.

NO.ACC

Gross and net weights cannot be

accumulated together

Only accumulate gross or net

weights (only works in LFT)

Battery fails to charge

fully

Battery is defective

Have battery replaced by Ohaus

authorized service dealer.