Rice Lake SURVIVOR OTR Steel Deck User Manual

Page 31

Advertising
background image

Trimming and Calibration

27

1. Park the cart as close as possible to side 1 being trimmed with the wheels centered over the end load cell

mount (Number 1 in Figure 7-2). Record the indicator reading. Remember that the scale is still
uncalibrated, so the indicator readings are simply raw counts rather than weight units.

Figure 8-2. Trimming Load Cell Number One

2. Move the cart directly over mount Number 2 and record that reading. Move the cart directly over mount

Number 3 and record that reading. Move the cart so the wheels are centered directly over mount Number 4
(you may have to turn the weight cart around so all wheels remain on the scale) and record the reading

Figure 8-3. Trimming Load Cell Numbers Two, Three, and Four

3. The lowest reading of the four is the reference cell. Do not change that cell’s signal. Instead use the

individual cell potentiometers for the other three cells to reduce those signals to match your reference cell.
Remember that you turned all pots to full signal before starting, so you can not increase the signal from any
cell; only decrease signal by trimming with the pots.

4. Note that the best trim is always the least trim. If one of the four readings differs from the others by more

than 5% of the displayed counts, there is probably a mechanical problem with that load cell mount causing
the large difference. Find it and correct it before going on. Check for binding, an out-of-level or misaligned
link, or similar problems with the load cell and mount. Do not try to trim down large signal differences
with resistance pots - that only adds larger problems later because of interaction between mounts.

5. Park the loaded weight cart over one of the high-reading cells on Side 1. Turn that cell’s individual

potentiometer until the displayed reading equals your recorded reference cell reading. Repeat for the other
two high-reading cells on Side 1.

6. As adjustments are somewhat interactive, repeat the process in steps 1-5 until all four cells on Side 1 read

within .1% of each other.

7. Repeat steps 1-6 for load cells 5-8 on Side 2 of the scale.

1

4

3

2

8

5

6

7

1

4

3

2

8

5

6

7