4. The scale will perform a tare and the message “LOAD” will be displayed.

5. When the message “LOAD” is displayed, place the calibration weight on the scale’s hanging pan.

Calibration will begin automatically and “WAIT” will be displayed indicating that calibration is in
progress.



6. When finished the scale will automatically return to the weighing mode, and the calibration weight can be

removed from the pan.