Chapter 34: maintenance, Chapter 33: common errors and troubleshooting – Torbal BAS Series User Manual

Page 57

Advertising
background image

- 56 -

Error or Indicator

Cause

Explanation / Solution

- - - -

Below zero

Re-zero the scale by pressing the
Ж0Е key.

- -

Taring is not allowed

Place a weight on the pan before
taking a tare.

- -

Re-zeroing is not allowed

Remove weight from pan and clear
stored tare values.

L

Pan error

Make sure that pan is properly
seated on the pan support.

H

Exceeded capacity

The scale has exceeded its
weighing capacity. Reduce the
weight.

Err - b

Pan not cleared on power-up

The pan was not empty while the
scale was initiating at startup.
Clear the pan and restart the scale.

Sapl LO (Parts Counting)

Average piece weight is too low

The average piece weight is lower
than 3e. Be sure that individual
piece weight is greater than 3e.

Sapl LO (Percent Weighing)

Sample reference weight is too low The sample reference weight is

lower than 100d. Increase the
sample weight.

The scale will not turn on

Possible power failure

Check the power AC adapter
connection.

Weighing results are not accurate Inaccurate calibration

Calibrate the scale.

Weighing result will not stabilize Irregular environment

Eliminate drafts or vibrations.





Cleaning and maintaining your Scale:

• Before cleaning the scale always unplug the A/C adapter from the electrical outlet.

• Use a soft, slightly damp cloth to clean the exterior housing of your scale,.

• Wipe the scale gently. Do not allow any liquid to enter into the scale.
• Do not apply extensive pressure to the LCD or the LED display.

• Do not use chemicals or benzene when cleaning the surface. Corrosive chemicals may damage the

finish.

• Alcohol may be used only to clean the scale’s stainless steel pan or the draft ring.


Chapter 34: Maintenance

Chapter 33: Common Errors and Troubleshooting