5 Putting into operation

22

NIRMaster Operation Manual, Version B

Basic installation conditions

!

DANGER

Death or serious injuries by use in explosive environments.
• Do not store or operate the instrument in explosive environments
• Remove all sources of flammable vapors
• Do not store chemicals in the vicinity of the device

The instrument must be installed:
• indoor only with a clearance of 15 cm minimum to walls and any other objects
• without being exposed to heating or cooling source (e.g. direct sunlight or air conditioning)
• with the transportation lock in unlocked position (see section 9.2)
• on a stable and horizontal position

If the instrument is installed on a laboratory cart, the cart should have large, soft wheels (approx. 10
cm diameter) for better shock absorption at transport on even floors.

Worst case environmental conditions at installation site must meet IP class and all other tech-
nical data (see section 3 ff .) .

5 .3

Electrical connection requirements

To maintain the systems ingress protection against liquids, dust and others, a proper installation
is required! Therefore, installation of cables (e .g . for a touchscreen monitor) must be performed
by a service technician with appropriate training only .

After the pre-installation procedure (all internal connections have been established and transportation
lock is in unlocked position) has been completed successfully, the power plug of the NIRMaster can
be connected to mains.

The used mains circuit has to:
• provide the voltage that is given on the type plate of the instrument.
• be able to handle the load of the connected instruments.
• be equipped with adequat fuses and electrical safety measures, in particular proper grounding.

See also technical data of all components regarding the different minimum system requirements!

NOTICE

Risk of instrument damage by wrong mains supply.
• External mains supply must meet the voltage given on the type plate
• Check for proper grounding
• Exchange defective cabling instantly