8 .3 reference measurement fails – BUCHI NIRMaste Pro IP65 User Manual

Page 59

Advertising
background image

8 Troubleshooting

59

NIRMaster Operation Manual, Version B

8 .3

Reference measurement fails

When the deviation between the stored reference spectrum is out of limit, the reference measurement
will fail. A pop-up window will show both curves and give details of the measurement result.

Possible reasons for a reference measurement to fail
• Optical surfaces (e.g. External reference) dirty or scratched
• Environmental conditions out of operational range (e.g. condensating humidity at optical surfaces)
• Defective instrument (e.g. electronic damage)
• After lamp-module exchange

Remedy
• Clean the External reference and the NIRMaster output window
• Place the External reference onto the output window and press the ‘Rerun’ button to repeat the

measurement

If measurement still fails, a new reference spectrum can be stored by pressing the ‘Checkmark’
button. A login window will pop-up. An administrator login is required to store the new data set!