8 Troubleshooting

60

NIRMaster Operation Manual, Version B

Buttons ‘External Reference Deviation’

Button

Functionality

‘Exit’

‘Rerun’

‘Checkmark’

NOTE
After lamp-module exchange, a reference spectrum should be generated to match the individual light
intensity output of the module. In all other cases, perform consequent trouble shooting (run SST)
before setting a new reference spectrum!

8 .4

Connection to a local network

The embedded PC is equipped with two Ethernet (LAN) sockets.

2

1

LAN 2
The secondary network connector (LAN2) can be used to connect the NIRMaster to a local Ethernet
network.

By default, LAN 2 is configured to obtain an IP address automatically (via DHCP). A customer specific
static IP-address and subnet mask for LAN 2 can be configured if required by a local network admin-
istrator with administrative rights on the NIRMaster.

LAN 1
The primary network connector (LAN1) is predefined for internal communication with the spectrom-
eter hardware.

Standard configuration of

LAN 1

:

• Static IP-address 192.168.1.2
• Subnet mask 255.255.255.0

Standard configuration of the Spectrometer:
• Static IP-address 192.168.1.1
• Subnet mask 255.255.255.0

NOTE
Do NOT change the settings between the embedded PC and spectrometer!