BUCHI NIRMaste Pro IP65 User Manual

Page 23

Advertising
background image

5 Putting into operation

23

NIRMaster Operation Manual, Version B

NOTE
• To cut the power in case of an emergency by unplugging, the instrument or any other item must

not block the mains plug! In this case, the plug must be able to be pulled out instantly.

• When the system is installed with a locked or secured plug an external power switch is required.

• When the system is installed stationary (without power plug) an external power switch is

required.

• External power switches (e.g. emergency stop switches) must meet IEC 60947-1 and IEC

60947-3 requirements. Such devices must be clearly labeled and accessible at any time.

• Additional electrical safety measures such as residual current breakers may be necessary to meet

local laws and regulations!

• External connections and extension lines must be provided with a grounded conductor lead

(3-pole couplings, cord or plug equipment). All used power cords must meet the input power
requirements.