BUCHI NIRMaste Pro IP65 User Manual

Page 54

Advertising
background image

7 Maintenance and repairs

54

NIRMaster Operation Manual, Version B

7 .6 .2

Lamp exchange

The light source of the NIRMaster has a life expectancy of approx. 6000 hours. After this time, it
needs to be replaced. To avoid downtime, the NIRMaster is equipped with an independent secondary
backup lamp.

In case of a primary lamp module failure:
• The system automatically switches to the secondary lamp in case of a primary lamp failure.
• The secondary lamp will be in use until the primary lamp has been exchanged by the user or a

service person.

• The operator will be reminded to exchange the primary lamp module. This should be done as soon

as possible.

!

CAUTION

Risk of minor burns by hot lamp module.
• Do not touch hot parts or surfaces
• Let the lamp module cool down safely
• Do not touch the bulb glass

a

d

c

e

b

Advertising