7 Maintenance and repairs

53

NIRMaster Operation Manual, Version B

7 .6

NIRMaster basic instrument

The drawing below shows the main components of the NIRMaster at open housing.

Interferometer

Primary lamp

Interferometer
interface panel

LASER assembly

ON / OFF button

Cable feedthrough

USB Connector

Internal fuses

Add-on Connector

7 .6 .1

Resetting the system fuses

The primary fuses are located inside the instrument next to the interferometer power supply socket. To
get access to the resettable fuses, lift off the top cover as described in section.

Fuse triggered

Fuse ok

F1 — F2 (4 A, 250V)

Resetting a system fuse
• Push in the fuse for reset
• Switch ON the NIRMaster

NOTE
Device fuses may be triggered occasionally (e.g. due to voltage peaks). When fuses trigger
frequently, inform your customer service! Do not try to repair or modify the fuses!