70

Bath Unit connections

For information regarding Bath Unit connections, refer to the CAV
Installation & Setup Guide.

SDU-200 Solvent Dispensing System

The CAV is equipped with a solvent system used to wash the viscometer
tubes after each sample test. The system consists of the Solvent Dispens-
ing Unit (SDU), the internal and external drain lines, and the user-
supplied solvent and waste cans. Appropriate solvent containers may be
acquired from many suppliers, including JustRite

®

and Nalgene

®

(see

weblinks below).

http://www.justritemfg.com/containes.tpl
http://www.nalgenelabware.com/products/productsIndex.asp

The SDU will interface with the CANNON Service Unit (CSU) and the
CAV instrument.

Solvent system connections

For information regarding solvent system connections, refer to the CAV
Installation & Setup Guide.

CSU-200 Service Unit

The Service Unit (CSU-200) contains control electronics, vacuum pump
and pressure regulators for air, solvent and vacuum. Each CSU-200 is
capable of providing regulated gas/vacuum for up to four Bath Units.

NOTE

If the user has in-house regulation of gas and vacuum, these connec-
tions can be made directly to the Bath Unit. See chapter 5 for required
pressure/vacuum settings.

CSU-200 connections

For information regarding CSU-200 connections, refer to the CAV
Installation & Setup Guide.

NOTE

A compressed air source is required for connection to the Service Unit.

Emptying the waste can

CAUTION

Check the volume of the waste can at the beginning of every work shift,
to avoid solvent spillage or overflow.

The waste can collects solvent and sample residue drained from the
viscometer tubes during the cleaning process. CANNON

®

Instrument

Company does not supply a waste can. The CAV should be connected to
a suitable safe container, approved by local safety personnel. The drain
system operates on a gravity flow system.

CAUTION

The drain line must be kept short and neat (no loops or knots) or an air
lock could result.