Dakota Ultrasonics

98

9.3 Zero Coating

In order to account for very slight electronic differences in transducers of the same
type, frequency, and diameter, the DFX-7 has been equipped with a “zero coating”
feature. This enables the DFX-7 to obtain very accurate readings on coatings,
eliminating potential errors incurred from slight differences in the manufacturing
processes. The procedure is outlined below:

Performing a Coating Zero

1) Press the MULTI MODE key once to activate the measurement mode

options.

2) Use the UP and DOWN arrow keys to scroll through the sub menu items

until Coating Only (CT) mode is highlighted.

3) Press the ENTER key to select the measurement mode and return to the

measurement screen.

4) Apply a drop of couplant on the transducer and place the transducer in

steady contact with the probe zero disk (battery cover) and obtain a steady

reading.

Note: The coating measurement displayed will potentially be a value greater or

less than 0.