Checkline DFX-7 User Manual

Page 59

Advertising
background image

DFX-7

Ultrasonic Flaw Detector

55

ESC key multiple times to tab left until the CAL menu is highlighted and

displaying the submenu items.

3) Use the UP and DOWN arrow keys to scroll through the sub menu items

until MATL 1PT is highlighted.

4) Press the ENTER key to display the Digits Edit Box.

5) Press the UP and DOWN arrow keys to scroll the highlighted value.

6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.

8) Press the OK key to calculate the velocity and return to the menu screen, or

ESC to cancel the one point calibration.

9) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.

Note: CHECK YOUR CALIBRATION! Place the transducer back on the

calibration point. The thickness reading should now match the known

thickness. If the thickness is not correct, repeat the steps above.

At some point there may become a requirement for improved accuracy over a smaller
measurement range. In this case, a two point calibration would be most suited for