Checkline DFX-7 User Manual

Page 9

Advertising
background image

DFX-7

Ultrasonic Flaw Detector

5

2.2 Auto Probe Recognition

When the DFX-7 is initially powered up, the gauge will automatically check to see if
the transducer plugged into the gauge can be recognized. The steps that follow
assume the DFX-7 recognized the probe type:

Probe Automatically Recognized

1) Press the OK key once to use the identified probe, or ESC to display a list of

optional transducers. Note: if the

DFX-7

recognizes a specific transducer,

the user should always select OK to use the identified probe. The only time

an alternative probe should be selected from a list is if the user switched

probes following initial power up and recognition.

2) Assuming the

DFX-7

recognized the probe and the OK key was pressed,

the

DFX-7

will advance to a Zero Probe menu. If the transducer was

identified as a special transducer capable of measuring coating thickness, a

menu will be displayed allowing the user the ability to toggle the coating

thickness display on/off as follows: