Checkline DFX-7 User Manual

Page 132

Advertising
background image

Dakota Ultrasonics

128

material. There are three width options (SPIKE, THIN, and WIDE). The SPIKE
setting may be desirable for high resolution and general applications to decrease the
overall noise. This can be considered the normal or standard setting. When
additional energy is needed, more penetration, the THIN and WIDE options may be
necessary. The procedure to change the pulse width is outlined below:

Selecting the Pulse Width

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the TUNE menu is highlighted and displaying the submenu items.

2) Use the UP and DOWN arrow keys to scroll through the sub menu items

until PULSE is highlighted.

3) Use the LEFT and RIGHT arrow keys to scroll the Pulse options.

4) Once the Pulse is displayed, press the MEAS key to return to the

measurement screen.

10.13 Pulser Voltage

The DFX-7 has a 200 volt square wave pulser that can be adjusted for specific
applications and transducers. The Pulser Volt feature offers a 50 volt cut/boost

Advertising