Checkline DFX-7 User Manual

Page 109

Advertising
background image

DFX-7

Ultrasonic Flaw Detector

105

One Point Calibration

Note: Use the maximum coating sample for the one point calibration first.

1) Physically measure the thicker of the two samples of coating, as close as

possible to the maximum expected coating measurement range, using a set

of calipers or a digital micrometer.

Very Important:

If coating measurements will be made with the coating

applied to a metal surface, the calibration must be done in the same manner,

with the samples coupled to a metal surface. However, if the coating will be

measured as a stand alone material, the calibration must be performed the

same way.

2) Apply a drop of couplant on the transducer and place the transducer in

steady contact with the thicker of the two coating and samples fist. Be sure

that the reading is stable and the repeatability indicator, in the top left corner

of the display, is fully lit and stable. Press the MENU key once to activate

the menu items tab. Press the MENU key multiple times to tab right and the

ESC key multiple times to tab left until the CAL menu is highlighted and

displaying the submenu items.

Advertising