DFX-7

Ultrasonic Flaw Detector

127

Note: Before toggling the Polarity, the

DFX-7

should be set to the RF display

view option. The RF view will give the user the best opportunity to correctly

view the positive and negative cycles of the waveform. Please refer to page 61

for information on selecting the Display Views.

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right, and the ESC key multiple times to tab left,

until the TUNE menu is highlighted and displaying the submenu items.

2) Use the UP and DOWN arrow keys to scroll through the sub menu items

until POLARITY is highlighted.

3) Use the LEFT and RIGHT arrow keys to toggle the POLARITY negative or

positive.

4) Press the MEAS key to return to the measurement screen and begin taking

readings.

Important: Be sure to do a Probe Zero after changing the polarity!

10.12 Pulse Width

The DFX-7 has an adjustable pulse width option. Pulse width, refers to the duration
of time the pulser is left on. This time results in increased energy sent into the test