Checkline DFX-7 User Manual

Page 55

Advertising
background image

DFX-7

Ultrasonic Flaw Detector

51

Coating Probe Identified Coating Probe Not Identified

5) Press the ENTER key to display the confirmation screen.

6) If a coating transducer was identified use the UP and DOWN arrow keys to

toggle coating on/off.

7) Press the OK key to complete the probe zero function, or ESC key to cancel

the probe zero function.

8) Remove the transducer from the probe zero disk, and proceed to the

calibration section.

Note: The value that is displayed will change depending on the current velocity

setting in the CMX. Disregard the number that is displayed. It is not

important. What is important is accurately performing the steps outlined above

to insure reliability of the probe zero calculation.

6.3 Material Calibration

In order for the DFX-7 to make accurate measurements, it must be set to the correct
sound velocity of the material being measured. Different types of materials have
different inherent sound velocities. For example, the velocity of sound through steel
is about 0.233 inches per microsecond, versus that of aluminum, which is about
0.248 inches per microsecond. If the gauge is not set to the correct sound velocity,