Types of error and error code, Types of error and error, Error code list – KEYENCE DL-PN1 User Manual

Page 31

Advertising
background image

29

- PROFINET Communication Unit DL-PN1 User's Manual (IB) -

Types of error and error code

Error code list

The following error codes may occur in the DL-PN1.

DL-PN1

Error code

Description

Cause

Actions

0

No error

-

-

51

Unassigned ID error

The main unit assigned no ID
within 10 seconds after the
DL-PN1 had been started.

• Check if the number of connected sensor

amplifiers exceeds the maximum number
of sensor amplifiers that can be
connected to the main unit.

"Types and Number of Connectable

Sensor Amplifiers" (page 3)

• Check the connection with the sensor

amplifiers and then turn the power on
again.
If this error cannot be recovered, contact
your nearest KEYENCE office.

52

Start-time
communication error

Communication between sensor
amplifiers ended abnormally
before ID assignment
completion.

Check the connection with the sensor
amplifiers and then turn the power on again.
If this error cannot be recovered, contact
your nearest KEYENCE office.

53

Unsupported sensor
amplifier connection
error

A sensor amplifier not supported
by the DL-PN1 is connected.

Check the model of the connected sensor
amplifier and remove the sensor amplifier if
it is not supported by the DL-PN1.

54

Mixed model error

Sensor amplifiers outside the
specifications have a mixed
connection.

Check if the models are mixable.

55

Start-time
communication error

ID number assignment is
successful but communication
failed during the subsequent
initial communication.

Check the connection with the sensor
amplifiers and then turn the power on again.
If this error cannot be recovered, contact
your nearest KEYENCE office.

56

Current limitation
error

The number of connected
sensor amplifiers exceeds the
allowable range.

Use sensor amplifiers within the allowable
range.

57

Communication error
between sensor
amplifiers

An error occurred during
communication between sensor
amplifiers.

Check if there is a noise source around the
DL-PN1.
If the sensor communication indicator is
solid red, turn the power on again.

100

System error

The IP address is incorrect.

Contact your nearest KEYENCE office.

101

System error

A default gateway setting error
occurred.

102

System error

An attempt to read data in
EEPROM such as the MAC
address has failed.

103

System error

An attempt to start the protocol
stack has failed.

104

System error

An attempt to access FlashROM
has failed.

150

System error

The number of held IDs is
incorrect.

151

System error

The number of sensors is
incorrect.

152

System error

An initial read error occurred.

Advertising